journal article Aug 23, 2002

Correction of concomitant magnetic field‐induced image artifacts in nonaxial echo‐planar imaging

Magnetic Resonance in Medicine Vol. 48 No. 3 pp. 509-515 · Wiley
View at Publisher Save 10.1002/mrm.10249
Abstract
AbstractEcho‐planar images acquired in nonaxial planes are often distorted. Such image distortion has limited the applications of the echo‐planar imaging (EPI) technique. In this article, it is demonstrated that a considerable amount of the distortion is caused by the higher‐order magnetic field concomitant with the linear magnetic field gradient, or the concomitant magnetic field. The image distortion caused by the concomitant magnetic field is more prominent when a higher gradient amplitude is used for readout. It is also shown that the concomitant magnetic field can cause ghosting and blurring. A theoretical analysis is performed for the concomitant field effect in nonaxial EPI images. A point‐by‐point (or line‐by‐line) phase correction algorithm is developed to correct the image distortion, ghosting, and blurring. A postreconstruction processing algorithm is also developed to correct image distortion with much higher computational efficiency. Experimental results show that both correction methods effectively reduce the image distortion in coronal or sagittal images. Magn Reson Med 48:509–515, 2002. © 2002 Wiley‐Liss, Inc.
Topics

No keywords indexed for this article. Browse by subject →

References
16
[2]
Norris DG (1985)
[3]
Coxon R (1989)
[6]
Correction for geometric distortion in echo planar images from B0 field variations

Peter Jezzard, Robert S. Balaban

Magnetic Resonance in Medicine 10.1002/mrm.1910340111
[12]
Bracewell RN (1978)
[14]
GloverGH PelcNJ. Method for correction of image distortion due to gradient non‐uniformity. US Patent 4 591 789;1986.
Metrics
62
Citations
16
References
Details
Published
Aug 23, 2002
Vol/Issue
48(3)
Pages
509-515
License
View
Cite This Article
Yiping P. Du, Xiaohong Joe Zhou, Matt A. Bernstein (2002). Correction of concomitant magnetic field‐induced image artifacts in nonaxial echo‐planar imaging. Magnetic Resonance in Medicine, 48(3), 509-515. https://doi.org/10.1002/mrm.10249
Related

You May Also Like

Sparse MRI: The application of compressed sensing for rapid MR imaging

Michael Lustig, David Donoho · 2007

5,679 citations

Generalized autocalibrating partially parallel acquisitions (GRAPPA)

Mark A. Griswold, Peter M. Jakob · 2002

4,656 citations

SENSE: Sensitivity encoding for fast MRI

Klaas P. Pruessmann, Markus Weiger · 1999

4,133 citations

Movement‐Related effects in fMRI time‐series

Karl J. Friston, Steven Williams · 1996

3,445 citations