Robust computer‐controlled system for intracytoplasmic sperm injection and subsequent cell electro‐activation

Abstract
AbstractIntroductionIntracytoplasmic sperm injection (ICSI) and the subsequent cell electro‐activation process is a relatively new enhanced procedure to address male factor infertility. The current method involves the engagement of experienced embryologists for such a purpose. More advanced methodologies, which use high precision instrumentation tools, will speed up the whole procedure.MethodsIn this paper, the development of a computer‐controlled system for ICSI and the subsequent cell electro‐activation process is presented. The system is integrated to a microinjection workstation and piezo‐actuator to perform the ICSI procedure, with vision capability to automatically position the components precisely. A micro‐pump assembly is utilized for automatic medium refreshment and a heater plate assembly provides temperature control during the cell electro‐activation process. The overall system is comprehensive, comprising modular functional components integrated within a hardware architecture.ResultsExperimental results on mice oocytes verified the effectiveness of the developed system over the current method.ConclusionsFurther improvements on the instrumentation tools will improve the robustness and overall performance of the developed system. Copyright © 2009 John Wiley & Sons, Ltd.
Topics

No keywords indexed for this article. Browse by subject →

References
20
[6]
Oh YK "Ultrastructure of rabbit ova recovered from ovarian follicles and inseminated in vitro" Fertil Steril (1975)
[11]
Whitacker MJ "Calcium and cell cycle control" Development (1990)
[13]
HiguchiT KudohKI MimatsuA. Microinjection apparatus and method of controlling microinjection. US Patent No. 5225750 1993.
[14]
HiguchiT WatanabeM. Apparatus for effecting fine movement by impact force produced by piezoelectric or electrostrictive element. US Patent No. 4894579 1990.
[15]
HiguchiT KudohKI MimatsuA. Microdrive Apparatus. US Patent No. 5229679 1993.
[17]
dSPACE GmbH. dSPACE DS1102 User's Guide. Document Version 3. dSPACE: Paderborn Germany 1996.
[18]
Shapiro L (2001)
[20]
TanKK NgSC XieY.Optimal intracytoplasmic sperm injection with a piezo micromanipulator. In Proceedings of the 4th World Congress on Intelligent Control and Automation2002;4(1)1120–1123. 10.1109/wcica.2002.1020753
Metrics
3
Citations
20
References
Details
Published
Jan 23, 2009
Vol/Issue
5(1)
Pages
85-98
License
View
Cite This Article
K. K. Tan, S. Huang, K. Z. Tang (2009). Robust computer‐controlled system for intracytoplasmic sperm injection and subsequent cell electro‐activation. The International Journal of Medical Robotics and Computer Assisted Surgery, 5(1), 85-98. https://doi.org/10.1002/rcs.239