Impact of the Positive Pressure Valve on Vascular Access Device Occlusions and Bloodstream Infections

View at Publisher Save 10.2309/java.14-2-6
Abstract
Abstract

Background/Purpose: The use of needleless positive pressure connector valves (PPVs) on venous access devices (VADs) has been shown to have possible benefits in decreasing VAD occlusion rates. Catheter-related bloodstream infection (CRBSI) rates, however, have been variable. With this potential for improving patient outcomes, decreasing costs, and facilitating nursing care and catheter management, the use of PPV devices becomes both advantageous and desirable.

Methodology: This was a randomized, prospective, parallel clinical study of 160 medical and surgical step-down unit patients requiring parenteral therapies. Patients were equally randomized to two study groups to assess the impact on occlusion and CRBSI rates, using a PPV versus a standard cap without PPV. VADs included both peripherally inserted catheters (PICCs) and midlines.

Results: Seven VAD occlusions occurred; six in the experimental group, one in the control group (p = 0.43). Also, two CRBSIs occurred in the experimental group, with none in the control group (p = 0.497). All incidents occurred in doublelumen PICCs. The lack of statistical difference between the two groups for occlusions and CRBSIs indicates that the use of PPVs had no impact in preventing such outcomes. However, it was observed that the number of double-lumen PICCs and their dwelling time in the experimental group was significantly higher (p = 0.003).

Conclusion: There was no difference in the rate of catheter occlusions or CRBSIs when using PPVs or standard caps. We consider that the study outcomes may be due to the study's small sample size, and we speculate that longer dwell time of double lumen catheters may have contributed to these outcomes. These observations deserve further investigation. However, the use of PPVs may still be advantageous from a nursing process perspective.
Topics

No keywords indexed for this article. Browse by subject →

References
26
[1]
Berger "The effects of positive pressure devices on catheter occlusions" Journal of Vascular Access Devices Winter (2000)
[2]
Bouza "A needleless closed system device (CLAVE) protects from intravascular catheter tip and hub colonization: a prospective randomized study" Journal of Hospital Infection (2003) 10.1016/s0195-6701(03)00136-1
[3]
Bowers "Comparison of occlusion rates by flushing solutions for peripherally inserted central catheters with positive pressure lueractivated devices" Journal of Infusion Nursing (2008) 10.1097/01.nan.0000308542.90615.c2
[4]
Casey "A randomized, prospective clinical trial to assess the potential infection risk associated with the Posiflow™ needleless connector" Journal of Hospital Infection (2003) 10.1016/s0195-6701(03)00130-0
[5]
Crnick "The promise of novel technology for the prevention of intravascular device-related bloodstream infection. I. Pathogenesis and short-term devices" Clinical Infectious Diseases (2002) 10.1086/339863
[6]
Elliott "Effects of heparin and chlorbutol on bacterial colonisation of intravascular cannulae in an in vitro model" Journal of Hospital Infection (1989) 10.1016/0195-6701(89)90035-2
[7]
Esteve "Bloodstream infection related to catheter connections: a prospective trial of two connection systems" Journal of Hospital Infection (2007) 10.1016/j.jhin.2007.05.021
[8]
Field "Incidence of catheter-related bloodstream infection among patients with a needleless, mechanical valve-based intravenous connector in an Australian hematology-oncology unit" Infection Control and Hospital Epidemiology (2007) 10.1086/516660
[9]
Gil "Triple- vs. single-lumen central venous catheters. A prospective study in a critically ill population" Archives of Internal Medicine (1989) 10.1001/archinte.1989.00390050105021
[10]
Hamilton "Complications associated with venous access devices: part two" Nursing Standard (2006) 10.7748/ns2006.03.20.27.59.c4095
[11]
Hoch "Management of the complications of long-term venous access" Seminars in Vascular Surgery (1997)
[12]
Jacobs "Central venous catheter occlusion: a prospective, controlled trial examining the impact of a positive pressure valve device" JPEN Journal of Parenteral and Enteral Nutrition (2004) 10.1177/0148607104028002113
[13]
Jasinsky "Occlusion reduction and heparin elimination using an anti-reflux device on peripheral and central venous catheters" Journal of Infusion Nursing (2009) 10.1097/nan.0b013e3181921c56
[14]
Knue "The efficacy and safety of blood sampling through peripherally inserted central catheter devices in children" Journal of Infusion Nursing (2005) 10.1097/00129804-200501000-00004
[15]
Krzywda "Predisposing factors, prevention, and management of central venous catheter occlusions" Journal of Intravenous Nursing (1999)
[16]
Maragakis "Increased catheter-related bloodstream infection rates after the introduction of a new mechanical valve intravenous access port" Infection Control and Hospital Epidemiology (2006) 10.1086/499166
[17]
O'Grady "Centers for Disease Control and Prevention. Guidelines for the prevention of intravascular catheter-related infections" MMWR Recommendations and Reports (2002)
[18]
Rummel "Clinical evaluation of a positive pressure device to prevent central venous catheter occlusion: results of a pilot study" Clinical Journal of Oncology Nursing (2001)
[19]
Outbreak of Bloodstream Infection Temporally Associated with the Use of an Intravascular Needleless Valve

M. E. Rupp, L. A. Sholtz, D. R. Jourdan et al.

Clinical Infectious Diseases 2007 10.1086/517538
[20]
Safdar "A review of risk factors for catheter-related bloodstream infection caused by percutaneously inserted, noncuffed central venous catheters: implications for preventive strategies" Medicine (2002) 10.1097/00005792-200211000-00007
[21]
Salgado "Increased rate of catheter-related bloodstream infection associated with use of a needleless mechanical valve device at a long-term acute care hospital" Infection Control and Hospital Epidemiology (2007) 10.1086/516800
[22]
Schilling "The impact of needleless connector device design on central venous catheter occlusion in children: a prospective, controlled trial" JPEN Journal of Parenteral and Enteral Nutrition (2006) 10.1177/014860710603000285
[23]
Stephens "Are clinical signs accurate indicators of the cause of central venous catheter occlusion?" JPEN Journal of Parenteral and Enteral Nutrition (1995) 10.1177/014860719501900175
[24]
Yebenes "Clinical use of disinfectable needle-free connectors" American Journal of Infection Control (2008) 10.1016/j.ajic.2008.10.013
[25]
Yebenes "Prevention of catheter-related bloodstream infection in critically ill patients using a disinfectable, needle-free connector: a randomized controlled trial" American Journal of Infection Control (2004) 10.1016/j.ajic.2003.12.004
[26]
Yeung "Infection rate for single-lumen vs. triple-lumen subclavian catheters" Infection Control and Hospital Epidemiology 1988 (1988) 10.1086/645820
Metrics
21
Citations
26
References
Details
Published
Jan 01, 2009
Vol/Issue
14(2)
Pages
84-91
License
View
Cite This Article
Nabil Khalidi, Debra S. Kovacevich, Lorelei F. Papke-O'Donnell, et al. (2009). Impact of the Positive Pressure Valve on Vascular Access Device Occlusions and Bloodstream Infections. Journal of the Association for Vascular Access, 14(2), 84-91. https://doi.org/10.2309/java.14-2-6