Abstract
ABSTRACT

Objectives
To measure the thickness homogeneity of Invisalign (Align Technology Inc, San José, Calif) aligners with micro-computed tomography (micro-CT) scans.


Materials and Methods
Starting from micro-CT scanning of 20 different aligners, multiplanar reconstructions were obtained. An orthodontist blinded about the study measured aligner thickness in different regions (molar, canine, incisor) and in different sites (gingival–buccal, buccal, occlusal, lingual, and gingival–lingual). To assess various thicknesses in different aligner sites and regions, the sample was stratified into subgroups and linear regression analysis was performed.


Results
Descriptive analysis showed that mean thickness of aligners in the incisor region ranged from 0.582 mm to 0.639 mm, in the canine region from 0.569 mm to 0.644 mm, and in the molar region from 0.566 mm to 0.634 mm. Student's t-tests showed no significant differences in the aligner thickness of different regions when data were stratified by different sites. Student's t-tests showed significant differences in thickness homogeneity for the molar region when the data were stratified by tooth (mean difference = 0.068 mm; 95% confidence interval, 0.009–0.126 mm; P = .024).


Conclusions
Invisalign aligner thickness is characterized by small differences. The only significant difference was revealed in the molar region where thickness of the gingival–lingual edge is significantly thinner than that measured at the occlusal aspect. From a clinical perspective, the results of the present study could be considered to explain the reduced predictability of several orthodontic tooth movements in the molar region.
Topics

No keywords indexed for this article. Browse by subject →

References
36
[1]
Flores-Mir CBJ, Pacheco-Pereira C. Patient satisfaction and quality of life status after 2 treatment modalities: Invisalign and conventional fixed appliances. Am J Orthod Dentofac Orthop. 2018; 154(5): 639– 644. 10.1016/j.ajodo.2018.01.013
[2]
Rosvall MD, Fields HW, Ziuchkovski J, Rosenstiel SF, Johnston WM. Attractiveness, acceptability, and value of orthodontic appliances. Am J Orthod Dentofac Orthop. 2009; 135(3): 276– 277. 10.1016/j.ajodo.2008.09.020
[3]
White DW, Julien KC, Jacob H, Campbell PM, Buschang PH. Discomfort associated with Invisalign and traditional brackets: a randomized, prospective trial. Angle Orthod. 2017; 87(6): 801– 808. 10.2319/091416-687.1
[4]
Meazzini MC, Mazzoleni F, Autelitano L. Comparison of pain perception in patients affected by cleft and cranio facial anomalies treated with traditional fixed appliances or Invisalign. Cleft Palate Craniofac J. 2020; 57(1): 35– 42. 10.1177/1055665619861780
[5]
Rossini G, Parrini S, Castroflorio T, Deregibus A, Debernardi CL. Periodontal health during clear aligners treatment: a systematic review. Eur J Orthod. 2015; 37(5): 539– 543. 10.1093/ejo/cju083
[6]
Buschang PH, Chastain D, Keylor CL, Crosby D, Julien KC. Incidence of white spot lesions among patients treated with clear aligners and traditional braces. Angle Orthod. 2019; 89(3): 359– 364. 10.2319/073118-553.1
[7]
Efficacy of clear aligners in controlling orthodontic tooth movement: A systematic review

Gabriele Rossini, Simone Parrini, Tommaso Castroflorio et al.

The Angle Orthodontist 10.2319/061614-436.1
[8]
Effectiveness of clear aligner therapy for orthodontic treatment: A systematic review

Lindsay Robertson, Harsimrat Kaur, Nathalia Carolina Fernandes Fagundes et al.

Orthodontics & Craniofacial Research 10.1111/ocr.12353
[9]
Clear aligner orthodontic therapy of rotated mandibular round-shaped teeth: A finite element study

Andrea Cortona, Gabriele Rossini, Simone Parrini et al.

The Angle Orthodontist 10.2319/020719-86.1
[10]
Grunheid T, Loh C, Larson BE. How accurate is Invisalign in nonextraction cases? Are predicted tooth positions achieved? Angle Orthod. 2017; 87(6): 809– 815. 10.2319/022717-147.1
[11]
Garino FCT, Daher S, Ravera S, Rossini G, Cugliari G, Deregibus A. Effectiveness of composite attachments in controlling upper-molar movement with aligners. J Clin Orthod. 2016; 50(6): 341– 347.
[12]
Hahn WDH, Fialka-Fricke J, Fricke-Zech S, Zapf A, Kubein-Meesenburg D, Sadat-Khonsari R. Influence of thermoplastic appliance thickness on the magnitude of force delivered to a maxillary central incisor during tipping. Am J Orthod Dentofac Orthop.2009; 136(1): 12, e11–e17; discussion 12–13.
[13]
Kohda N, Iijima M, Muguruma T, Brantley WA, Ahluwalia KS, Mizoguchi I. Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances. Angle Orthod. 2013; 83(3): 476– 483. 10.2319/052512-432.1
[14]
Papageorgiou S, Koletsi D, Iliadi A, Peltomäki T, Eliades T. Treatment outcome with orthodontic aligners and fixed appliances: a systematic review with meta-analyses. Eur J Orthod. 2020; 42(3): 331– 343. 10.1093/ejo/cjz094
[15]
Lombardo L, Palone M, Longo M, et al. MicroCT X-ray comparison of aligner gap and thickness of six brands of aligners: an in-vitro study. Prog Orthod. 2020; 21(1): 12. 10.1186/s40510-020-00312-w
[16]
Gao L, Wichelhaus A. Forces and moments delivered by the PET-G aligner to a maxillary central incisor for palatal tipping and intrusion. Angle Orthod. 2017; 87(4): 534– 541. 10.2319/090216-666.1
[17]
Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine

F. Elkholy, B. Mikhaiel, F. Schmidt et al.

Journal of Orofacial Orthopedics / Fortschritte de... 10.1007/s00056-017-0090-4
[18]
Elkholy F. Recommendation of a novel film-thickness sequence, 0.4, 0.5 and 0.75 mm, for aligner systems. J Align Orthod. 2018(4): 295– 304.
[19]
Bucci RRR, Levate C, Michelotti A, Barone S, Razionale AV, D'Anto V. Thickness of orthodontic clear aligners after thermoforming and after 10 days of intraoral exposure: a prospective clinical study. Prog Orthod. 2019; 20(1): 36. 10.1186/s40510-019-0289-6
[20]
Ryu JH, Kwon JS, Jiang HB, Cha JY, Kim KM. Effects of thermoforming on the physical and mechanical properties of thermoplastic materials for transparent orthodontic aligners. Kor J Orthod. 2018; 48(5): 316– 325. 10.4041/kjod.2018.48.5.316
[21]
Kwon JS, Lee YK, Lim BS, Lim YK. Force delivery properties of thermoplastic orthodontic materials. Am J Orthod Dentofac Orthop. 2008; 133(2): 228– 234. 10.1016/j.ajodo.2006.03.034
[22]
Scanning electron microscopy evaluation of aligner fit on teeth

Edoardo Mantovani, Enrico Castroflorio, Gabriele Rossini et al.

The Angle Orthodontist 10.2319/120417-827.1
[23]
Mantovani E, Castroflorio E, Rossini G, et al. Scanning electron microscopy analysis of aligner fitting on anchorage attachments. J Orofac Orthop. 2019; 80(2): 79– 87. 10.1007/s00056-018-00167-1
[24]
Chow SC, Shao J, Wang H. Sample size calculations in clinical research. 2nd ed. Boca Raton: Chapman & Hall/CRC;2008.
[25]
Galvão MCSSJ, Coelho EC. Dahlberg formula—a novel approach for its evaluation. Dental Press J Orthod. 2012; 17(1): 115– 124. 10.1590/s2176-94512012000100015
[26]
Eksi O, Karabeyoğlu S. The effect of process parameters on thickness distribution in thermoforming. Adv Sci Technol Res J. 2017; 11(2): 198– 204. 10.12913/22998624/71147
[27]
Condo RPL, Cerroni L, Pasquantonio G, et al. Mechanical properties of “two generations” of teeth aligners: change analysis during oral permanence. Dent Mater J. 2018; 37(5): 835– 842. 10.4012/dmj.2017-323
[28]
Martina S, Rongo R, Bucci R, Razionale AV, Valletta R, D'Anto V. In vitro cytotoxicity of different thermoplastic materials for clear aligners. Angle Orthod. 2019; 89(6): 942– 945. 10.2319/091718-674.1
[29]
Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine

F. Elkholy, B. Mikhaiel, F. Schmidt et al.

Journal of Orofacial Orthopedics / Fortschritte de... 10.1007/s00056-017-0090-4
[30]
Elkholy F, Schmidt F, Jager R, Lapatki BG. Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: an in-vitro study. Am J Orthod Dentofac Orthop. 2017; 151(2): 407– 415. 10.1016/j.ajodo.2016.08.020
[31]
Iliadi A, Koletsi D, Eliades T. Forces and moments generated by aligner-type appliances for orthodontic tooth movement: a systematic review and meta-analysis. Orthod Craniofac Res. 2019; 22(4): 248– 258. 10.1111/ocr.12333
[32]
Krey K-F. Behaviour of monolayer and multilayer foils in the aligner thermoforming process. J Align Orthod. 2019; 3(2): 139– 145.
[33]
Houle JP, Piedade L, Todescan R Jr, Pinheiro FH. The predictability of transverse changes with Invisalign. Angle Orthod. 2017; 87(1): 19– 24. 10.2319/122115-875.1
[34]
Cattaneo PMDM, Melsen B. The transfer of occlusal forces through the maxillary molars: a finite element study. Am J Orthod Dentofac Orthop. 2003; 123(4): 367– 373. 10.1067/mod.2003.73
[35]
Garnett BSMK, Nguyen M, Al-Khateeb A, Liu S, Boyd R, Oh H. Cephalometric comparison of adult anterior open bite treatment using clear aligners and fixed appliances. Angle Orthod. 2019; 89(1): 3– 9. 10.2319/010418-4.1
[36]
Papadimitriou A, Mousoulea S, Gkantidis N, Kloukos D. Clinical effectiveness of Invisalign orthodontic treatment: a systematic review. Prog Orthod. 2018; 19(1): 37. 10.1186/s40510-018-0235-z