journal article Jul 24, 2018

Toward in vivo dosimetry in external beam radiotherapy using x‐ray acoustic computed tomography: A soft‐tissue phantom study validation

Medical Physics Vol. 45 No. 9 pp. 4191-4200 · Wiley
Abstract
PurposeTo study, using phantoms made from biological tissues, the feasibility and practical challenges of monitoring the position of the radiation beam and the deposited dose by x‐ray acoustic computed tomography (XACT) during external beam radiotherapy delivery.Material and methodsA prototype XACT system with a single immersion ultrasound transducer, which was positioned around the target sample driven by a motor‐controlled rotation stage, was used to acquire the x‐ray acoustic (XA) signals produced by a medical linear accelerator (Linac) to form an XACT image of the irradiated phantom. To investigate the feasibility of XACT in tracking the position of radiation dose, a large piece of veal liver with embedded fat tissue was imaged and beam misalignments were measured. Next, we explored the sensitivity of XACT in monitoring and quantifying the delivered dose, in which a block of porcine gel was embedded with equally spaced lard cylinders and imaged. The doses on the lard cylinders modulated by physical wedges were quantified from the XACT image and were verified by comparison to measurements from radiochromic films as the gold standard. Then, to simulate how XACT can perform in a targeted tissue in the human body, a porcine gel phantom with lard cylinders covered by different materials (bone, muscle, and air gap, respectively) was also imaged.ResultsThe reconstructed XACT images of the phantoms show congruence with the boundaries of the beam field and the interfaces between the different tissue materials. The beam displacement from the target was tracked properly by the reconstructed XACT images. An intensity difference as small as 2.9% in delivered dose region can be measured from XACT images P = 0.02. The intensities of XACT images were highly correlated to the film measurements with an R2 better than 0.986. The expected variances of dose delivered to different target regions as a result of the difference in attenuation were successfully captured by the XACT images.ConclusionsThis study validated the feasibility of XACT in accurately obtaining relative dose maps of tissue‐mimicking phantoms. XACT offers a practical method for verifying the beam position against the target and quantifying the relative dose delivered to the target during external beam radiotherapy.
Topics

No keywords indexed for this article. Browse by subject →

References
38
[1]
HalperinEC Wazer DE Perez CA BradyLW.Perez and Brady's principles and practice of radiation oncology 7th edn.Philadelphia PA:Wolters Kluwer;2018.
[2]
Podgorsak EB (2003)
[3]
Podgoršak EB (2006)
[4]
Hall EJ (2012)
[5]
Lehnert S (2008)
[8]
Sterzing F "Image‐guided radiotherapy" Deutsch Aerztebl Int (2011)
[9]
Timmerman RD (2012)
[12]
Huyskens DP (2001)
[13]
YorkeE AlecuR DingL et al.Diode in vivo dosimetry for patients receiving external beam radiation therapy.Report of Task Group;2005 62. 10.37206/88
[19]
Photoacoustic Tomography: In Vivo Imaging from Organelles to Organs

Lihong V. Wang, Song Hu

Science 10.1126/science.1216210
[21]
Experimental evaluation of x‐ray acoustic computed tomography for radiotherapy dosimetry applications

Susannah Hickling, Hao Lei, Maritza Hobson et al.

Medical Physics 10.1002/mp.12039
[24]
X‐ray‐induced acoustic computed tomography for 3D breast imaging: A simulation study

Shanshan Tang, Kai Yang, Liangzhong Xiang

Medical Physics 10.1002/mp.12829
[25]
Gunderson LL (2015)
[33]
CoxBT LauferJG BeardPC.The challenges for quantitative photoacoustic imaging. Paper presented at: SPIE BiOS;2009. 10.1117/12.806788
[35]
HubbellJH SeltzerSM.Tables of X‐ray mass attenuation coefficients and mass energy‐absorption coefficients 1 keV to 20 MeV for elements Z = 1 to 92 and 48 additional substances of dosimetric interest. National Inst. of Standards and Technology‐PL Gaithersburg MD (United States). Ionizing Radiation Div.;1995. 10.6028/nist.ir.5632
[38]
X-ray characterisation of normal and neoplastic breast tissues

P C Johns, M J Yaffe

Physics in Medicine and Biology 10.1088/0031-9155/32/6/002
Cited By
38